Using 50-Ohm Coax From DUT to Oscilloscope

What if we remove the probes from the equation by connecting our DUT's output directly to the oscilloscope's analog input using a 3-ft. 50-Ω coaxial cable?

Using 50-Ohm Coax From DUT to Oscilloscope

In our recent exploration of 10x passive probes, we've determined that while these types of probes are great general-purpose tools, they're not necessarily going to do the job in specialized measurement circumstances. They're relatively low-bandwidth, low-SNR probes that impose some limitations and, in some scenarios, can deliver potentially misleading or erroneous measurement results if used without clear understanding of their capabilities.

But what if we remove the probes from the equation by connecting our DUT's output directly to the oscilloscope's analog input using a 3-ft. 50-Ω coaxial cable? What if we also use a 1-MΩ input impedance on our oscilloscope when doing so?

Figure 1: A coaxial cable presents high impedance at low frequencies but acts as a transmission line at higher frequencies.

As Figure 1 reveals, a 50-Ω coaxial cable exhibits high impedance at low frequencies, but there's the input impedance imposed by the cable's capacitance to consider. At low frequency, the cable looks like a capacitor of roughly 100 pF (30 pF/foot for RG-58 coax). That's about 10x the capacitive load of the 10x passive probe.

If you're using the oscilloscope's 1-MΩ input, at low frequencies this will look like a 100-pF load. As frequency increases to, say, the 10-MHz range, we'll see some low impedances in the range of 10-100 Ω. That can load down your DUT considerably. It's like adding a 100-pF capacitor to your circuit.

Figure 2: A low-impedance source will result in multiple reflections and ringing.

At higher frequencies, the coax will look more like a transmission line, and the signal will bounce back and forth with reflections. With the oscilloscope input set for 1-MΩ impedance and depending on the impedance of the DUT, ringing is a possibility.

This is a scenario where situational awareness is critical. We have to know the source impedance of the DUT to understand the impact of reflections. If the source impedance driving the cable is 50 Ω, there's no problem. We'll get half the voltage into the transmission line, it travels down the line, sees the high impedance at the oscilloscope input, bounces back, and is terminated by the source impedance.

But rarely will our DUT present a 50-Ω source impedance to the transmission line. If it's a very low-impedance source, such as a power rail, we'll see a lot of reflection between the source and the 1-MΩ impedance of the oscilloscope input (Figure 2).

If it's a high-impedance source, we won't get much voltage launched into the line. The signal will reflect back and forth and gradually build up with a long rise time, looking like an RC filter charging up (Figure 3), even though it isn't one.

Figure 3: A high-impedance source will look like an RC filter charging up with a long, slow rise time.

If we're working with frequencies in the range of 20-50 MHz and signals at volt levels, we're just as well served, if not better, by simply using our 10x passive probe, which won't load down the DUT nearly as much as the coaxial cable will. At, say, 10-20 MHz, the cable is usable as long as we bear in mind that it presents a 100-pF load as noted above.

If we want high bandwidth and we want to use a 50-Ω cable, then we need to terminate the cable with a 50-Ω impedance at the oscilloscope input. We can achieve very high bandwidths this way. However, we cannot use AC coupling in the oscilloscope. For high DC impedance, we'll need to use an external coaxial DC-blocking capacitor. But, if we use that blocking capacitor and we're using the 50-Ω input to eliminate reflections, we've made a high-pass filter and we will not be able to see the low-frequency components of the signal.

Figure 4: The addition of a tiny amount of tip inductance severely impacts the transfer function of a coaxial connection.

Our bandwidth is still limited by the tip inductance and cable attenuation. One cannot overemphasize how important the tip inductance is. For example, in Figure 4, we see the attenuation of RG-174 coaxial cable alone with good coaxial connections (in red), and we also see the response that we get if we add a small breakout region at the tip (in blue), as if we'd added a coax connector soldered to the board. By simply pulling the signal and return apart and adding perhaps 9 or 10 nH or loop inductance, we reap a dramatic drop in the transfer function.

At the end of the day, if we want to connect our DUT to an oscilloscope using only coaxial cable and coaxial connections and also maximize bandwidth, we want the cleanest interface possible with the least amount of impedance discontinuity.

Copyright © 2020-2026 Teledyne LeCroy. All rights reserved. All original content, including text and photos, is the property of Teledyne LeCroy and cannot be reproduced without expressed written permission.

Editor's Note: There are two important considerations when connecting a 50 Ohm cable directly to a 50 Ohm terminated oscilloscope input channel: 1) The max input voltage for an oscilloscope 50 Ohm terminated input is usually very low and your power rail voltage may exceed this maximum input rating and damage your oscilloscope input front-end amplifier. 2) For maximum usability, you will want your oscilloscope to have very high input channel offset so that you can offset the DC rail voltage to the vertical middle of the oscilloscope display grid (0 volts) and then use a high-sensitivity setting (e.g. 1, 2, 5 or 10 mV/div) to view small variations. Many oscilloscopes do not offer enough offset at high-sensitivity settings to permit this type of operation (Teledyne LeCroy HDOs are notable exceptions).

Eric Bogatin presented a webinar Secrets to Successful Power Rail Measurements. Click the button below to watch this webinar, or download the transcript to preview the webinar content. Additional related webinar material from Eric Bogatin on power rail and power integrity measurements is included in this same link.

An 8-part webinar series on Power Integrity may also be of interest to readers. Click the button below to learn more about this webinar series.

Great! You’ve successfully signed up.

Welcome back! You've successfully signed in.

You've successfully subscribed to Teledyne LeCroy Oscilloscope Blog.

Success! Check your email for magic link to sign-in.

Success! Your billing info has been updated.

Your billing was not updated.